ECHOGRAPH 1090 BASIC with all necessary functions for ultrasonic testing

ECHOGRAPH 1090 DAC
with additional features for an easy signal evaluation with DAC

ECHOGRAPH 1090 DGS/DAC
with the additional possibility of a comfortable DGS-evaluation
  
   Compact, sturdy, digital - makes Ultrasonic Testing easier

The new digital and light weight ultrasonic instrument ECHOGRAPH 1090 makes ultrasonic testing much easier. 

This is the result of 
  • a small housing
  • a large colour display (143 x 79 mmІ) for A-scan display, measured results and control menu which is good readable also from distances.
  • clearly arranged keys for the main functions at the front panel and
  • the avoidance of abbreviations in the menu.
An additional feature is the new wizard function control for guidance of ultrasonic operators with little experience through the main instrument settings and the probe calibration.
The instrument with an aluminium housing is an absolute light weight instrument with only 1.6 kg including lithium ion battery for an operating time of 13 hours.
The thickness of the ECHOGRAPH 1090 housing is only 50 mm and it includes very fast digital ultrasonic electronic components with a pulse repetition frequency up to 1500 Hz, two gates to monitor the ultrasonic time of flight and amplitude, a large A-scan memory and two control lights at the front side assigned to the corresponding gates.
The ECHOGRAPH 1090 is consequently the perfect instrument for manual ultrasonic testing. The coloured rubber frame protects the instrument against impact damage and rough surface conditions and guaranties a stable standing.
The interface offers the possibility to use the ECHOGRAPH 1090 as an instrument in a small automatic testing system. This is due to the monitor signal outputs at the interface and the statistical processing of the signals.
 
   Featured highlights (amongst others) ...
  
DGS/DAC version

DGS and DAC curve with 4 additional evaluation curves direct
on display and
amplitude evaluation relative to DGS/DAC-curve (dBrel)

TFT-LC-Display with 6.3" (16 cm) A-screen size suitable for daylight, can be read in direct sunlight, 256 colours
Easy operating by plain language and windows technique
Five freely programmable function keys
Important instrument parameters accessible via direct keys
  
   Plus...
  

compact aluminium case with protective rubber holster for
harsh ambient conditions

built-in Li-Ion battery for up to 13 hrs operating time, safe automatic quick charge function controlled by internal charging processor

energy saving in battery mode

selectable colour display for measuring mode and menu operation

operating language, parameter and help texts editable via PC and
can be stored in the instrument*

assistant for various instrument settings, e.g. for determination
of the delay line of the probe, when adjusting the screen
display, DAC and DGS settings

Reference A-Scan in the background

echo-dynamic curve

USB interface

pulse repetition frequency 8 Hz to 1500 Hz

monitor real-time output for both monitor gates

trigger modes: internal, external (in/out), 1st echo

update and/or upgrade of the operating software via PC
(CD-ROM, e-mail, download)*

manufactured and tested for reliability according to EN 12668-1


* available on request

   Order Information

ECHOGRAPH 1090 Basic
incl. red protective rubber holster, Li-Ion rechargeable battery, mains/charger unit and transport case

1090.101

ECHOGRAPH 1090 DAC
incl. red protective rubber holster, Li-Ion rechargeable battery, mains/charger unit and transport case
1090.201
ECHOGRAPH 1090 DGS/DAC
incl. red protective rubber holster, Li-Ion rechargeable battery, mains/charger unit and transport case
1090.301
  
Accessories 
Carrying case incl. carrying strap 6189.001
Carrying strap 6189.002
PC link cable (USB) 1657.702
 
Software
ECOM 90,
PC-software for Windows XP/2000
1995.005
 
Options
Protective holster (black) 6189.006
Protective holster (yellow) 6189.007
Protective holster (green) 6189.008
 

   ECHOGRAPH 1090 Technical Data Overlook

Screen
Screen type Colour LC display
transmissive / transreflective
suitable for daylight
background illumination
Screen size 143.4 x 79.3 mmІ
Resolution 400 x 240 pix, 256 colours
A-scan size 142 x 73.5 mmІ
Scale electronically generated, can be switched on/off
Scale division coarse: 10 sections horizontally, 5 sections vertically
fine: 50 sections horizontally, 25 sections vertically


A-scan representation and digitizing
Image repetition frequency 50 Hz
A-scan representation envelope
inverse (filled)
freeze
echo-dynamic curve
zoom across gate 1
RF representation across the entire adjustment range
Rectification full-wave, positive, negative, RF
Suppression selectable manually: 0 - 99% screen height in steps of 1% (linear)
Zoom gate range (gate 1) to full screen width
A/D converter 9 bit
Digitizing process direct, w/ A/D converters
Scan rate 80 MHz
Scan error during digitising < +/- 0.5% screen height at 4 MHz
Reaction delay < 20 ms
 
Sweep range
Time base range 2.5 - 4850 mm steel
Sound velocity 100 - 15000 m/s in steps of 1 m/s
Delayed time base sweep 0 - 3000 mm in steps of 0.1 mm
Linearity of time base +/- 0.5 % of the screen width
pulse repetition frequency 8 Hz to 1500 Hz (depending on time base range, selectable from -85% to +50% in steps of 1%)
Trigger modes internally, externally, 1st echo
 
Transmitters
Number of transmitters 2 (1 resolution, 1 power)
Shape of TX pulse Unidirectional (negative) needle pulse
TX damping 10, 50, 220, w/o [Ohm]
 
Amplifier and damping
Number of frequency ranges 3 (LF- and RF range, broadband)
Adjustable damping 100 dB in steps of 0.1 - 1 - 2 - 6 - 12 - 20 dB


Echo evaluation, determination of flaw size
Indication of echo height
(valid for both gates)
in % screen height (FSH)
dBrel (DAC version)
dBabs (DAC version)
mmFBH (DGS/DAC-Version)
Indication of echo run time sound path
depth and projection distance and/or reduced projection distance (with angled scan)
resolution 0.1 mmSt


Distance Amplitude Correction (DAC version)
Number of points max. 11
DAC-curve shift Anhebung/Absenkung: max. ± 80 dB
Additional evaluation curves max. 4 (up to ± 15 dB to DAC-curve)
 
DGS method (DGS/DAC version)
DGS curve 0 t 30 mmFBH snd backwall
Reference reflector backwall or FBH
Additional evaluation curves max. 4 (up to ± 15 dB to DGS-curve)
 
Monitor gates
Number of monitor gates 2
Response time Pulse repetition frequency (max. 1500 Hz)
Operating modes normal, inverted, off
Adjustment range Gate start:
0 - 3000 mm in steps of 0.1 mm
Gate width:
0 - 1000 mm in steps of 0.1 mm
Statistical clearing (both monitor gates) 0 - 250 occurrences
Switching outputs go / no go (both monitor gates) Level: TTL (5V) , low active, ZA = 100 Ohm
Response accuracy: +/- 0.5% screen height
Switching hysteresis: < 0.5% screen height
hold time of the switched output: 12 ms
Optical indication (both monitor gates) 2 LED's on the front panel


Inputs and outputs
USB interface USB1 interface for PC connection and for printing via PC
VGA output* for an external monitor
Synchronizing input/output* TTL - level (5V), low active,
trigger threshold 2 V approx.
trigger in/out is standard fitting


Miscellaneous
Measuring systems selectable mm, inch
Date and time Built-in real-time clock
Languages German, English, one additional language* can be stored in the instrument via PC (the texts are editable by means of a PC)


Storage facilities
A-scan representation Current A-scan representation on the screen by means of the freeze key
Data sets 224 data sets incl. A-scans, parameters, time and text* (8 lines with 37 characters each) in the internal memory


Power supply
Mains operation via mains unit 1808.501:
85 - 264 VAC, 47 - 63 Hz
Output: 10 VDC
Operating temperature: 0 °C - +50 °C
Storage temperature: -40 °C - +85 °C
Permissible humidity: 0 - 95%
Operating time w/ battery supply (with full load) With built-in Li-Ion battery
7 h approx. when the illumination is switched on
13 h approx. when the illumination is switched off
Energy saving modus on/off
Automatic switch-off with undervoltage of mains or battery supply
 
Mechanics
Size (H x W x D) 166 mm x 201 mm x 50 mm w/o protective frame
190 mm x 217 mm x 64 mm w/ protective frame
Front panel (H x W) 147 x 178 mmІ
Weight 2.0 kg (with Li-Ion battery and protective holster)
1.6 kg (with Li-Ion battery and w/o protective holster)
Connectors 2 x Lemo 1 (for probes)
PC: USB 1
D-type sub (9 pin) for service, flaw output and external trigger
connection of mains unit: 5.5 mm DC plug

* available on request